QS5Vacuum PumpOPERATING MANUAL
8 QS5 Operating Manual9.0 Trouble-ShootingProblem Possible Cause RemedyDoes not achieve good vacuum1. Loose/open fittings2. Not enough oil3. Dirty/c
9QS5 Operating Manual10.0 WarrantyFor technical and customer assistance, please call 1-800-344-3304. Do not return pump to INFICON. Contact your l
TWO TECHNOLOGY PLACEEAST SYRACUSE, NY 13057-9714 USAPhone: 1-800-344-3304Fax: 315-434-2552E-Mail: [email protected]
Preface QS5 Operating ManualThank you for buying the INFICON QS5 Vacuum Pump! QS5 is a high-quality, two-stage rotary vane vacuum pump designed for fa
1QS5 Operating ManualTable Of Contents1.0 The QS5’s High Performance Features. . . . . . . . . . . . . 22.0 Specifications . . . . . . . . . . . . . .
2 QS5 Operating Manual1.0 The QS5’s High Performance Features High vacuum, high speed pumpTwo-stage rotary-vane design provides deep, 15 micron ult
3QS5 Operating Manual2.0 Specifications3.0 Safety PrecautionsRead this manual and become familiar with the specifications of this machine prior to
4 QS5 Operating ManualThe exhaust gas from the vacuum pump is composed of oil and gases from within the pump. The exhaust gas will be hot and should b
5QS5 Operating Manual5.0 Start-up ProcedureTHE QS5 IS NOT SHIPPED WITH OIL IN ITS RESERVOIR. DO NOT START PUMP WITHOUT ADDING OIL! USE ONLY HIGH VIS
6 QS5 Operating ManualIT IS IMPORTANT THAT THE PUMP HAS REACHED ITS NORMAL OPERATIONING TEMPERATURE (50°C-70°C / 122°F-158°F) BEFORE ANY CONDENSABLE V
7QS5 Operating Manual8.0 MaintenanceDISCONNECT QS5 FROM POWER SUPPLY BEFORE CARRYING OUT ANY MAINTENANCE.8.1 Changing the Oil THE OIL MUST BE CHANG